Evaluation of the Flash effect in breast irradiation using TomoDirect: an investigational study

This study assessed the theoretical dose distribution resulting from Flash in breast irradiation using TomoDirect. A cylindrical phantom that enabled dose distribution of the breast was used for verifying the effect of planning target volume (PTV) contouring and Flash. A total of 18 Gy in 10 fractions were prescribed to the PTV. Five PTVs were then created by Contracting this contour by 0, 1, 2, 3, 4 and 5 mm, giving PTV-x. Flash ±x is defined by opening x (number) of the leaves. The Flash effect in the air was compared with each set-up error of 5, 10 and 15 mm, respectively. The minimum PTV dose from PTV-1 to PTV-3 increased from 13.88 Gy to 15.86 Gy. In contrast, Dmin in PTV-4 and PTV-5 was 17.80 Gy in 98.88% of the prescription dose. Without Flash, when 5-, 10- and 15-mm set-up errors applied in the PTV, relative doses of 87.88, 23.73 and 7.94% were observed, respectively. However, in Flash 3, which was equal to the usual air margin of 1.875 cm, a relative dose of 104.24% ± 0.30% was observed, irrespective of set-up errors (5 mm to 15 mm). Flash opening is useful for countervailing set-up errors in breast cancer patients who receive breast irradiation with TomoDirect.
Source: Journal of Radiation Research - Category: Physics Authors: Tags: Physics Source Type: research