Implantation of a Novel Telemetric Intraocular Pressure Sensor in Patients With Glaucoma (ARGOS Study): 1-Year Results [Clinical Trials]

Conclusions. Despite early postoperative anterior chamber inflammation, the IOP sensor was well tolerated by all patients. We describe the first prospective clinical study of a noncontact IOP sensor that potentially enables continuous IOP monitoring in patients with glaucoma. The sensor shape and size needs to be adapted to avoid pupillary distortion and to confirm that IOP measurements are accurately recorded in comparison to those of GAT. (www.germanctr.de; number DRKS00003335.)
Source: Investigative Ophthalmology - Category: Opthalmology Authors: Tags: Clinical Trials Source Type: research