Does an iPad fixation disparity test give equivalent results to the Mallett near fixation disparity test?

ConclusionAlthough the measured differences between the instruments are not significant in a Wilcoxon analysis, a Bland & Altman approach shows them to be in some cases clinically unacceptable, therefore the instruments should not be used interchangeably. Previous research indicates that the Mallett unit performs reasonably well at detecting symptomatic individuals and determining a prismatic correction that is likely to be helpful. Further research is required to determine the performance of the iPad test in these functions and to assess the reproducibility of both instruments.ResumenAntecedentesSe han desarrollado diversos instrumentos para medir el prisma de alineamiento, que es el prisma que elimina la disparidad de fijación (heteroforia asociada). Entre estos instrumentos se incluyen la unidad de visión de cerca de Mallett y la recientemente desarrollada Thomson Vision Toolbox en el iPad. Como no existe investigación previa acerca del acuerdo entre estos instrumentos, los clínicos pueden cuestionarse si pueden usarse ambos tests de forma intercambiable.MétodosSe realizó una prueba de visión de cerca a 80 participantes, utilizando la unidad de Mallett y la prueba iPad de disparidad de fijación. Los datos fueron analizados de cuatro modos, para evaluar el acuerdo entre ambos instrumentos.ResultadosMuchos participantes reportaron ausencia de disparidad de fijación (horizontalmente 46,25%, verticalmente 82,5%), o prisma de alineamiento no significativo (horizontalme...