Electromechanical Model-Based Design and Testing of Fiber Scanners for Endoscopy

Forward-viewing catheters and scopes for diagnosing disease and guiding interventions in small ducts (less than 3  mm diameter) require wide-field high-quality imaging since scope tip bending is difficult and ineffective. A high-fidelity electromechanically coupled finite element (FE) model of a piezoelectric actuated resonant fiber scanner is presented, which enables improvement on the general design of fibe r-optic scanner geometry to increase scan frequency and field of view (FOV). Using the proposed model, parametric sweeps on the specific design variables achieved by acid etching of glass fiber are analyzed to identify their effect on scanner performance and to choose improved designs. The resulting complex fiber scanner design requires development of unique microfabrication techniques. Comparison of three model simulations and their experimental testing show that our proposed coupled model has prediction error of ≤12% with respect to experimental data, while other uncoupled models have up t o 39% error. The model and microfabrication techniques presented in this paper have significance for fiber scanning-based systems in that they demonstrate reliability for model-driven design and also flexibility for fiber scanner design of complex geometries, allowing for improvement on medical imag ing performance.
Source: Journal of Medical Devices, Transactions of the ASME - Category: Medical Devices Source Type: research