Measuring the Sensitivity of Total Knee Replacement Kinematics and Laxity to Soft Tissue Imbalances

Ligament balancing during total knee replacement (TKR) is receiving increased attention due to its influence on resulting joint kinematics and laxity. We employed a novel in vitro technique to measure the kinematics and laxity of TKR implants during gait, and measured how these characteristics are influenced by implant shape and soft tissue balancing, simulated using virtual ligaments. Compared with virtual ligaments that were equally balanced in flexion and extension, the largest changes in stance-phase tibiofemoral AP and IE kinematics occurred when the virtual ligaments were simulated to be tighter in extension (tibia offset 1.0 ± 0.1 mm posterior and 3.6 ± 0.1° externally rotated).
Source: Journal of Biomechanics - Category: Biomedical Science Authors: Source Type: research