Balloon assisted retrieval of broken catheter in a 10-months old baby with patent ductus arteriosus: Case report

Publication date: Available online 3 February 2018 Source:Journal of Indian College of Cardiology Author(s): Santosh Kumar Sinha, Mahmodula Razi, Puneet Aggarwal, Lokendra Rekwal, Anupam Singh, Nishant Kumar Abhishekh Percutaneous retrieval of intravascular foreign body can be done by snares, biopsy forceps, dormia basket or tip deflecting wires. Herein, we report percutaneous retrieval a fractured multipurpose diagnostic catheter by balloon assisted technique in a 10-months old baby while doing device closure of a patent ductus arteriosus. The catheter got fractured at the junction of shaft and the tip while manipulation to right ventricular outflow tract during antegrade wiring. The fractured proximal end was trapped at the tip of venous sheath which was successfully wired with percutaneous coronary angioplasty wire and retrieved with the help of 1.5×10mm and 2×10mm Maverick balloon (Boston scientific; USA). This is the first ever report of balloon assisted retrieval of intravascular foreign body.
Source: Journal of Indian College of Cardiology - Category: Cardiology Source Type: research