Failure to Treat Life-Threatening Ventricular Tachyarrhythmias in Contemporary Implantable Cardioverter-Defibrillators: Implications for Strategic Programming [Original Articles]

Conclusions: Complex and unanticipated interactions between manufacturer-specific features and generic programming can prevent therapy for VF. More data are needed to assess the risks and benefits of translating evidence-based detection parameters from one manufacturer to another.
Source: Circulation: Arrhythmia and Electrophysiology - Category: Cardiology Authors: Tags: Catheter Ablation and Implantable Cardioverter-Defibrillator, Sudden Cardiac Death, Ventricular Fibrillation Original Articles Source Type: research