Phase imaging quality improvement by modification of AFM probes ’ cantilever

In this study, the circularly shaped holes located in varying distance from the probe base, were cut out by focused ion beam. Such modification was a consequence of the results of the previous experiments (probe tip sharpening and cantilever thinning) where significant improvement of image quality in tapping and phase imaging mode has been revealed. The solution proposed herein gives similar results, but is much simpler from the technological point of view. Shorter exposition time of the tip onto gallium ions during FIB processing allows to reduce material degradation. The aim of this modification was to change harmonic oscillators’ properties in the simplest and fastest way, to obtain stronger signal for higher resonant frequencies, which can be advantageous for improving the quality of imaging in PI mode. Probes shaped in that way were used for AFM investigations with Bruker AFM nanoscope 8. As a testing material, titanium roughness standard sample, supplied by Bruker, was used. The results have shown that the modifications performed within these studies influence the oscillation of the probes, which in some cases may result in deterioration of the imaging quality under tapping mode for one or both self‐resonant frequencies. However, phase imaging results obtained using modified probes are of higher quality. The numerical simulations performed by application of finite element method were used to explain the results obtained experimentally. Phenomenon described within th...