Hemodynamic Profile of Two Aortic Endografts Accounting for Their Postimplantation Position

Endovascular aneurysm repair (EVAR) is a clinically effective technique for treating anatomically eligible abdominal aortic aneurysms (AAAs), involving the deployment of an endograft (EG) that is designed to prevent blood leakage in the aneurysmal sac. While most EGs have equivalent operating principles, the hemodynamic environment established by different EGs is not necessarily the same. So, to unveil the post-EVAR hemodynamic properties, we need an EG-specific computational approach that currently lacks from the literature. Endurant and Excluder are two EGs with similar pre-installation designs. We assumed that the flow conditions in the particular EGs do not vary significantly. The hypothesis was tested combining image reconstructions, computational fluid dynamics (CFD), and statistics, taking into account the postimplantation position of the EGs. Ten patients with Endurant EGs and ten patients with Excluder EGs were included in this study. The two groups were matched with respect to the preoperative morphological characteristics of the AAAs. The EG models are derived from image reconstructions of postoperative computed tomography scans. Wall shear stress (WSS), displacement force, velocity, and helicity were calculated in regions of interest within the EG structures, i.e., the main body, the upper and lower part of the limbs. Excluder generated higher WSS compared to Endurant, especially on the lower part of the limbs (p  = 0.001). Spatial fluctuations of WSS were obs...
Source: Journal of Medical Devices, Transactions of the ASME - Category: Medical Equipment Source Type: research