Design of Reverse Materials Resurfacing Implants for Mild –Moderate Medial Osteoarthritis of the Knee

The areas of the most frequent cartilage loss in mild –moderate medial osteoarthritis (OA) were reviewed from previous studies. Implant components were designed to resurface these areas. The surface geometries of the components were based on an average femur and tibia produced from 20 magnetic resonance imaging (MRI) models of normal knees. Accuracy of fit of the components was determined on these 20 individual knees. The femoral surface was toroidal, covering a band on the distal end of the femur, angled inward anteriorly. For a five-size system, the average deviations between the implant surfaces and the intact cartilage surfaces of 20 femurs were only 0.3 mm. For the tibia, the deviations were 0.5–0.7 mm, but the errors were mainly around the tibial spine, with smaller deviations in the central bearing region. Hence, these small implant components would accurately restore the original bearing surfaces and allow for preservation o f all the knee structures. Using a thin metal component for the tibia would preserve the strong cancellous bone near the surface, an advantage for fixation. In this case, the femoral component would have a plastic bearing surface, but still be less than 10 mm thickness. Such a design could have a useful place in the early treatment of medial OA of the knee.
Source: Journal of Medical Devices, Transactions of the ASME - Category: Medical Equipment Source Type: research