Eighty-six Percent Failure Rate of a Modular-Neck Femoral Stem Design at 3 to 5 Years: Lessons Learned

Conclusions: Emphasizing the reporting of positive results may leave orthopaedic surgeons reticent to publicize negative results; however, the high failure rate of this implant design within 5 years prompted this report. We believe that patients and orthopaedic surgeons should be made aware of this implant’s clinical problems and patients should be followed closely. Expedient revision is necessary when failure is identified, to minimize potentially severe tissue damage and metal toxicity. Level of Evidence: Therapeutic Level IV. See Instructions for Authors for a complete description of levels of evidence.
Source: JBJS - Category: Orthopaedics Authors: Tags: Hip The Orthopaedic Forum Source Type: research