Detection of 5-fluorouracil surface contamination in near real time

In this report, we describe the laboratory performance of a 5-fluorouracil (5-FU) monitor. Methods The monitor was evaluated by spiking ceramic, vinyl, composite, stainless steel, and glass surfaces of 100 cm2 area with 5-FU masses of 0, 5, 10, 25, 50, and 100 ng. The surface was sampled with a wetted cotton swab, the swab was extracted with buffer, and the resulting solution was applied to a lateral flow monitor. Two ways of evaluating the response of these monitors were used: an electronic method where a lateral flow reader was used for measuring line intensities, and a visual method where the intensity of the test line was visually compared to the control line. Results The 5-FU monitor is capable of detecting 10 ng/100 cm2 (0.1 ng/cm2) using the electronic reader and 25 ng/100 cm2 (0.25 ng/cm2) using the visual comparison method for the surfaces studied. The response of the monitors was compared to LC-MS/MS results for the same samples for validation and there was good correlation of the two methods but some differences in absolute response, especially at higher spiking levels for the surface samples.