Deep Bleeder Acoustic Coagulation (DBAC)—part II: in vivo testing of a research prototype system

Conclusions While not meeting all in vivo objectives, the overall performance of the DBAC applicator was positive. In particular, D&L automation workflow was verified during each of the tests, with processing times well under specified (10 min) limits, and all bleeder branches were detected and localized. Further, gross necropsy and tissue examination confirmed that the HIFU thermal lesions were coincident with the target vessel locations in over 90 % of the multi-array dosing treatments. The SFA/DFA bleeder models selected, and the protocols used, were the most suitable practical model options for the given DBAC anatomical and bleeder requirements. The animal models were imperfect in some challenging aspects, including requiring tissue-mimicking material (TMM) standoffs to achieve deep target depths, thereby introducing device-tissue motion, with resultant imaging artifacts. The model “bleeders” involved intact vessels, which are subject to less efficient heating and coagulation cascade behaviors than true puncture injuries.
Source: Journal of Therapeutic Ultrasound - Category: Radiology Source Type: research