Biocompatibility of silver containing silica films on Bioverit(R) II middle ear prostheses in rabbits

In this study, rabbits were implanted with Bioverit® II middle ear prostheses functionalized with silver containing dense and nanoporous silica films which were compared with pure silica coatings as well as silver sulfadiazine cream applied on nanoporous silica coating. The health status of animals was continuously monitored; blood was examined before and after implantation. After 21 days, the middle ears were inspected; implants and mucosal samples were processed for electron microscopy. Autopsies were performed and systemic spreading of silver was chemically analyzed exemplarily in liver and kidneys. For verification of direct cytotoxicity, NIH 3T3 cells were cultured on similar silver containing silica coatings on glass up to 3 days. In vitro a reduced viability of fibroblasts adhering directly on the samples was detected compared to cells growing on the surrounding plastic of the same culture dish. In transmission electron microscopy, phagocytosed silver silica fragments, silver sulfadiazine cream as well as silver nanoparticles were noticed inside endosomes. In vivo, clinical and post mortem examinations were inconspicuous. Chemical analyses showed no increased silver content compared to controls. Mucosal coverages on almost all prostheses were found. But reduction of granulation tissue was only obvious around silver-coated implants. Single necroses and apoptosis in the mucosa were correlated by intracellular accumulation of metallic silver. For confirming supportive...